We use cookies to ensure that we give you the best experience on our website. If you click 'Accept all cookies' we'll assume that you are happy to receive all cookies and you won't see this message again. If you click 'Reject all non-essential cookies' only necessary cookies providing core functionality such as security, network management, and accessibility will be enabled. Click 'Find out more' for information on how to change your cookie settings.

Aim: Paracetamol is a well-tolerated antipyretic widely used in severe malaria management. The study aimed to develop and validate a rapid LC–MS/MS assay to quantify paracetamol in plasma from patients with severe malaria. Materials & methods: Plasma sample was precipitated by organic solvent containing isotope-labeled paracetamol internal standard. Supernatant was isolated, diluted with water, followed by LC–MS/MS analysis. Results: Plasma samples were extracted and assayed in less than 5.5 min. The assay response was linear (0.125–50 mg/l) with total intra- and interassay imprecision of <1.4%, which were considerably lower than most published reports. Conclusion: We developed, validated and applied a rapid and small volume LC–MS/MS assay with high precision and accuracy for plasma paracetamol quantitation in 989 samples from 62 patients with severe malaria. The simple and high-throughput quality could facilitate assay automation for future clinical studies.